The HX200 is a high accuracy calibrated dew point or humidity transmitter. The remote sensor is designed to work at temperatures as high as 200°C (392°F). This high accuracy sensor is also built to withstand up to 750 psi. These transmitters are perfect for applications such as compressed air, dryers and environmental chambers.
Specifications Dew Point: -60 to 40°C (-76 to 104°F) Relative Humidity: 0 to 95% RH (non-condensing) unit displays dew point or humidity, as well as temperature Maximum Sensor Pressure Rating: 750 psi (51.7 bar) Sensor Body Material: 304 SS Maximum Operating Temperature: 200°C (392°F) (at 0% RH) Mesh Filter Top: Stainless Steel Single Output: 4 to 20 mA loop powered (relative humidity or dew point) Dew Point Accuracy: ±1.0°C from -20 to 40°C (-4 to 104°F); ±3.0°C below -20°C (-4°F) Relative Humidity Accuracy: ±1% RH from 15 to 40°C (59 to 104°F), 0 to 90% RH (non-condensing); ±2% RH below 15°C (59°F) or above 40°C (104°F) Display: 2 line LCD Resolution: 0.1 Power: 4 to 20 mA loop powered, 2-wire 24 Vdc (typical) 12 to 28 Vdc Dimensions (Enclosure): 88.9 L x 63.5 W mm (3.5 x 2.5") Mounting Center: 78.7 mm (3.1") Dimensions of Probe: Length: 10.16 cm (4") OD: 1.27 cm (0.5") ½ NPT Fitting Weight: 1.8 kg (4 lb)
Rated 5 out of
5 by
Dean Friesen from
Excellent product.Easy to incorporate in our system, operates flawlessly.
Date published: 2017-12-05
Rated 5 out of
5 by
humiditymeasurement from
can be connected to daqgot this for measuring inlet temperature and RH. can be connected to daq
Date published: 2022-09-12
Rated 5 out of
5 by
Copper from
works great in extreme environmentsThis instrument is more accurate than all of the other brands we have tried. It works great even through a high vibration environment where others fail
Date published: 2020-11-12
Rated 5 out of
5 by
rs1022 from
Works Great!Just finished installing yesterday, easy set up. Being used in Dew Point mode, and switched over via internal dip switched. This unit is loop powered, so it will require a 24vdc separate power source.
Date published: 2018-08-21
Rated 4 out of
5 by
Fibreman12345 from
Good productUsing 4 to monitor dew point in compressed air system. Using a rs232 to ethernet comverter to pull data into database. Works great.
Improvements to get 5 stars
1. Make the sensor replaceable by adding a plug
2. Replace power plug with a 5 pin m12 so you do not have to drill a hole in the housing to get serial communication out of unit.
The manual indicates serial communications are possible but does not show the pin out to connect a cable. Could such documentation be added to the manual?
Asked by: cppmonkey
It's been one year and this still hasn't been added to the manual. As well, it's very unclear if the power connection can also be used to receive the 4-20 mA signal. The manual is poorly written.
Answered by: tomM
Date published: 2022-10-14
What is the recommended calibration method and calibration frequency for this unit?
Asked by: AnhT
Hi AnhT
Thank you for the inquiry. Suggested calibration interval is 1 year.
Best regards
OMEGA
Answered by: NorwalkAE_RS
Date published: 2020-08-29
Can the HX200HR be adjusted if the measurements are off? e.g. ours is reading 77.9% at 80%
Asked by: calibratingjo
Thank you for the inquiry. Calibration is performed via pc connection at the factory. 27 points are stored in the processor. This item needs to be sent in for calibration.
Best regards
OMEGA
Answered by: NorwalkAE_RS
Date published: 2020-08-31
How is the RS-232 connection made? I would like to record the data from this sensor on a computer.
I also have a current input module on a DAQ board, could the data be recorded that way?
Asked by: Paul A
Hi Paul A,
Thank you for the inquiry. Manual shows the RS232 connection. But it will be easier to use the mA output with your DAQ system. With this family of sensor you have either RS232 or mA output but not both.
Best regards
OMEGA